Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Beckman DM10XL: (1 year) CAL VER /5520 DATE: 02-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 21 NUMBER OF LINES: 113 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: None 1.004 DISP Temperature: 0 - 50degC 1.004 DISP Humidity: <70% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {AC VOLTAGE} 1.007 DISP Connect the 5520A to the UUT as follows: 1.007 DISP [32] 5520A NORMAL HI to UUT V/OHM 1.007 DISP [32] 5520A NORMAL LO to UUT COM 1.008 HEAD AC VOLTAGE: {200V Range} 1.009 5520 200 190.0V 1% 0.4U 60H SI 2W 2.001 HEAD AC VOLTAGE: {750V Range} 2.002 5520 750 700V 1.5% 4U 60H SI 2W 3.001 HEAD {DC VOLTAGE} 3.002 HEAD DC VOLTAGE: {1000V Range} 3.003 5520 1000 1000V 0.7% 1U 2W 4.001 HEAD DC VOLTAGE: {200V Range} 4.002 5520 200 190.0V 0.7% 0.05/ 2W 5.001 HEAD DC VOLTAGE: {20V Range} 5.002 5520 20 19.00V 0.7% 0.05/ 2W 6.001 HEAD DC VOLTAGE: {2V Range} 6.002 5520 2 1.900V 0.7% 0.05/ 2W 7.001 HEAD DC VOLTAGE: {200mV Range} 7.002 5520 200 190.0mV 0.7% 0.05/ 2W 8.001 HEAD {RESISTANCE} 8.002 HEAD RESISTANCE: {20 MOhm Range} 8.003 5520 20 19.00MZ 3% 0.01U 2W 9.001 HEAD RESISTANCE: {2 MOhm Range} 9.002 5520 2 1.900MZ 1% 0.001U 2W 10.001 HEAD RESISTANCE: {200 kOhm Range} 10.002 5520 200 190.0kZ 1% 0.1U 2W 11.001 HEAD RESISTANCE: {20 kOhm Range} 11.002 DISP Connect the 5520A to the UUT as follows: 11.002 DISP [32] 5520A AUX HI to UUT V/OHM 11.002 DISP [32] 5520A AUX LO to UUT COM 11.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 11.002 DISP [32] 5520A NORMAL LO to UUT COM 11.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 11.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 11.003 5520 20 19.00kZ 1% 0.01U CW 12.001 HEAD RESISTANCE: {2 kOhm Range} 12.002 5520 2 1.900kZ 1% 0.001U CW 13.001 HEAD RESISTANCE: {200 Ohm Range} 13.002 5520 200 190.0Z 1% 0.3U CW 14.001 HEAD {DIODE TESTER} 14.002 PIC dm10xlv 14.003 DISP Set the DMM FUNCTION/RANGE switch to DIODE. 14.004 5520 0.5V S 2W 14.005 EVAL Does {the display read approx. 0.500 (diode junction)}? 15.001 HEAD {DC CURRENT} 15.002 DISP Connect the 5520A to the UUT as follows: 15.002 DISP [32] 5520A AUX HI to UUT 200mA 15.002 DISP [32] 5520A AUX LO to UUT COM 15.003 HEAD DC CURRENT: {200uA Range} 15.004 5520 200 190.0uA 1% 0.1U 2W 16.001 HEAD DC CURRENT: {2mA Range} 16.002 5520 2 1.900mA 1% 0.1U 2W 17.001 HEAD DC CURRENT: {20mA Range} 17.002 5520 20 19.00mA 1% 0.1U 2W 18.001 HEAD DC CURRENT: {200mA Range} 18.002 5520 200 190.0mA 1% 0.1U 2W 19.001 HEAD DC CURRENT: {10A Range} 19.002 DISP Connect the 5520A to the UUT as follows: 19.002 DISP [32] 5520A 20A to UUT 10A 19.002 DISP [32] 5520A AUX LO to UUT COM 19.003 5520 10 9.90A 2% 0.03U HC 2W 20.001 HEAD {SAFETY TESTER} 20.002 DISP Set the DMM FUNCTION/RANGE switch to SAFETY TESTER. 20.002 DISP 20.002 DISP *NOTE* this function does not operate off the internal 20.002 DISP [32] DMM battery, instead it is powered from the 20.002 DISP [32] voltage source being measured. The LCD display 20.002 DISP [32] will be BLANK. 20.003 DISP Connect the 5520A to the UUT as follows: 20.003 DISP [32] 5520A NORMAL HI to UUT V/OHM 20.003 DISP [32] 5520A NORMAL LO to UUT COM 20.004 5520 25V S 2W 20.005 EVAL Do {the LEDs indicate a voltage between 24 - 50}? 21.001 END